流式细胞术中应用的荧光染料介绍
2010-08-11 15:21阅读:
当我们在进行多色流式分析的时候,对分析成功的一个关键影响因素是每个抗体选择何种荧光染料标记。通常会有很多可行的结合,我们在做选择的时候有很多的因素需要我们去考虑。对任何单克隆抗体,其阴性和阳性的信噪比相差四到六倍都取决于荧光素的使用。相对的荧光强度也取决于使用的仪器。高密度表达的抗原我们可以应用任何荧光素标记的抗体来检测,低密度表达的抗原就需要我们应用高信噪比高的荧光素,譬如PE或者APC标记的抗体来充分检测分离出阳性表达的细胞与阴性表达的细胞。
自发荧光
个别的细胞有着他们各自特异性水平的自发荧光(荧光信号产生于他们自身)。当在全波长荧光通道中进行观察的时候,自发荧光信号在长波长(>600nm)明显的降低。
对于有较高水平自发荧光类型的细胞,如长期组织培养的细胞,应使用较高发射波长的荧光染料(APC、APC-Cy7等)标记抗体,通常他们会有一个较好信噪比的结果。
对于那些不是有较多自发荧光类型的细胞,如新鲜的细胞,可以使用FITC
标记的抗体了。
以下为各种常用荧光素介绍:
Alexa Fluor 488 has a spectrum almost identical to
that of fluorescein isothiocyanate (FITC), but with extraordinary
photostability. Because of this photostability, it has become a
choice for fluorescent microscopy applications and has become
popular in cytometry applications. It is detected in the FL1
detector of the FACSCalibur or FACScan. Unlike other fluorochromes
with similar emission spectra, A
lexa Fluor 488 is pH insensitive over a broad range.
Alexa Fluor 633 is a practical alternative to APC
as well as Cy5. Alexa Fluor 633 conjugates can be used in
multi-color flow cytometry with instruments equipped with a second
red laser or red diode. It is detected in the FL4 detector of the
FACSCalibur. The FACScan cannot be used to detect Alexa Fluor 633
conjugates because the FACScan lacks a red laser or diode. Like
other Alexa Fluor dyes, Alexa Fluor 633 exhibits uncommon
photostability, making it an ideal choice for fluorescent
microscopy.
A great number of different Alexa Fluor dyes exist that are beyond
the scope of this introductory fluorophore section. Many
manufacturers sell directly-conjugated Alexa Fluor antibodies. Know
that Molecular Probes' Zenon Antibody Labeling Kits, which are
available for all of their Alexa Fluor dyes, make it possible to
rapidly and quantitatively label antibodies from a purified
antibody fraction or from a crude antibody preparation such as
serum, ascites fluid or a hybridoma supernatant. See
http://www.probes.com/servlets/publications?id=150
for specifics.
Allophycocyanin (APC) is an accessory
photosynthetic pigment found in blue-green algae. APC has 6
phycocyanobilin chromophores per molecule, which are similar in
structure to phycoerythrobilin, the chromophore in phycoerythrin or
PE. APC tandem dyes, APC-Cy5.5 and APC-Cy7, are also available. APC
has a 650-nanometer wavelength absorption maximum and a
660-nanometer fluorescence emission maximum. APC can be used in
flow cytometers equipped with dual lasers for multi-color analysis.
Like Alexa Fluor 633, APC is excited using the helium-neon red
diode laser (633 nanometers) of the FACSCalibur and is detected on
the FL4 detector. APC cannot be detected on the FACScan as that
instrument is not equipped with a red laser.
APC-Cy7 is a tandem conjugate system that combines
APC and a cyanine dye (Cy7) and has an absorption maximum at ~650
nanometers. This tandem uses the efficiency of the fluorescence
light energy transfer between the two fluorochromes. When excited
by light from a helium-neon laser, the excited fluorochrome (APC)
is able to transfer its fluorescent energy to the cyanine molecule,
which then fluoresces at a longer wavelength. The resulting
fluorescent emission maximum is in the deep red at approximately
767 nanometers. APC-Cy7 run on a FACSCalibur results in dim
expression because the FL4 detector's optical filter is centered
for APC emission (660 nanometers) and not the longer red
wavelengths excited with the helium-neon diode. It is recommended
that special precautions be taken with this conjugate, and cells
stained with them, to protect the fluorochrome from long-term
exposure to visible light.
Carboxyfluorescein Diacetate (CFSE) can be used to
track asynchronous cell division. Cell division results in
sequential halving of the initial fluorescence, resulting in a
cellular fluorescence histogram.
The peaks labeled 1, 2, 3, 4 and 5 represent successive generations
of CFSE-cultured cells.
Cy3 and Cy5 are excited by the 488-nanometer line
of an argon laser and the 633-nanometer line of a helium-neon diode
or laser, respectively. These conjugates can be used in flow
cytometry but typically do not give the fluorescence intensity
comparable to that of PE or APC. Applications where a smaller dye
is required are more appropriate for these dyes. These
fluorochromes are well suited for fluorescent microscopy.
Enhanced Cyan Fluorescent Protein (eCFP) cannot be
analyzed using the FACScan or FACSCalibur as this molecule requires
excitation in the violet range. The protein is easily detected
using the violet laser of the LSRII, or the violet lines of the
MoFlo's argon or krypton lasers. This molecule has an excitation
maximum at 475 nanometers and is best excited using the MoFlo's
457nm line of the argon or argon-krypton mixed gas laser. Use of
the 407nm violet laser of the LSRII will result in weak eCFP
expression. More detailed discussion of this molecule can be found
in the next section of this web site, the section on Fluorescent
Proteins.
Enhanced Green Fluorescent Protein (eGFP) can be
excited at 488 nanometers with a peak emission at 509 nanometers
and is detected in the FL1 detector on the FACSCalibur or FACScan.
The MoFlo and LSRII are able to distinguish between concurrently
expressing eGFP and eYFP cells if the proper optical filters and
experimental controls exist. More detailed discussion of this
molecule can be found in the next section of this web site, the
section on Fluorescent Proteins.
Enhanced Yellow Fluorescent Protein (eYFP), a
yellow-shifted variant of the eGFP molecule, is also excited at 488
nanometers with a peak emission at 535 nanometers and is also
detected in the FL1 detector on the FACSCalibur or FACScan. More
detailed discussion of this molecule can be found in the next
section of this web site, the section on Fluorescent
Proteins.
Fluorescein isothiocyanate (FITC) is currently the
most commonly used fluorescent dye for flow cytometry analysis.
When excited at 488 nanometers, FITC has a green emission that's
usually collected at 530 nanometers, the FL1 detector of the
FACSCalibur or FACScan. FITC has a high quantum yield (efficiency
of energy transfer from absorption to emission fluorescence) and
approximately half of the absorbed photons are emitted as
fluorescent light. For fluorescent microscopy applications, FITC is
seldom used as it photobleaches rather quickly though in flow
cytometry applications, its photobleaching effects are not observed
due to a very brief interaction at the laser intercept. FITC is
highly sensitive to pH extremes.
Peridinin chlorophyll protein (PerCP) has a 677
nanometer maximum emission, red, when excited at 488 nanometers and
is detected on the FL3 detector of the FACSCalibur or FACScan. A
PerCP tandem dye is also available (PerCP-Cy5.5). PerCP is not
suited for the high-powered lasing (>150mW) applications, such
as on the MoFlo, due to its photobleaching characteristics. PerCP
conjugates can only be obtained from Becton Dickinson and its
subsidiary, Pharmingen.
Phycoerythrin (PE or R-PE) has a huge absorption
coefficient and almost perfect quantum efficiency. In vivo, it
functions to transfer light energy to chlorophyll during
photosynthesis. It is one of the brightest dyes used today and
emits in the yellow/orange at about 570 nanometers. Those
accustomed to fluorescent microscopy may not be familiar with this
fluorochrome as it photobleaches rather quickly under a
microscope.
Phycoerythrin-Cy5 (PE-Cy5) is a tandem conjugate
where PE is coupled to the cyan dye, Cy5. When excited by
488-nanometer light, the excited fluorochrome (PE) is able to
transfer its fluorescent energy to the cyanine molecule, which then
fluoresces at a longer wavelength in the red at 670 nanometers.
This tandem dye is known by a confusing myriad of names to include
Becton Dickinson's CyChrome, Caltag and Sigma's Tri-Color, GIBCO's
RED670, Coulter's PC5, and probably others. Other PE conjugates
exist, e.g., PE-Cy5.5 and PE-Cy7, that will not be discussed in
this introductory fluorophore section. It is recommended that
special precautions be taken with this conjugate, and cells stained
with them, to protect the fluorochrome from long-term exposure to
visible light.
Phycoerythrin-Texas Red (PE-Texas Red) is a tandem
conjugate where PE is coupled to Texas Red. Similar to other tandem
conjuates, when excited by 488-nanometer light, the excited
fluorochrome (PE) is able to transfer its fluorescent energy to the
Texas Red molecule, which then fluoresces at a longer wavelength
with a peak in the orange at 612 nanometers. This tandem is also
known by other names such as Red 612 and ECD (Electron Coupled
Dye). Know that PE-Texas Red conjugates run on the FACScan or
FACSCalibur will result in dull expression due to the extant
optical filters. This is not observed on the LSRII or MoFlo when
equipped with the appropriate optical filters for this conjugate.
There is considerable overlap of emission when running PE and
PE-Texas Red specimens.
Propidium Iodide (PI) is a membrane-impermeant dye
that stains by nondiscriminately intercalating into every 4th or
5th nucleic acid base pair, binding both DNA and RNA. Once bound,
PI undergoes a conformational change and becomes ~40 times
brighter. Propidium iodide has a broad emission spectrum with a
peak in the orange at 620 nanometers. A number of assays employ
propidium iodide. Cells or nuclei with altered DNA content are
identified by staining alcohol-fixed, RNAse-treated cells or nuclei
with this dye (see below).
The first peak (M1) on the left represents normal diploid cells.
The next major peak (M2) represents tumor cells with increased DNA
content and DNA index of 1.27. Propidium iodide can be combined
with additional fluorescent antibodies that are specific for a
unique cell population and allow for more accurate S phase analysis
of multiple overlapping populations.
Propidium iodide has also been employed for many
years as a marker for viability as the disrupted membranes of dead
cells allow the dye to pass freely to the nucleic acids. However,
this dye is very sticky; it will stick to sample tubing and, given
sufficient time exposure to living cells, living cells will appear
to be propidium iodide positive. Given this dye's broad emission
spectrum and its sticky properties, contemporary flow cytometry
labs have replaced propidium iodide with other nucleic acid dyes,
e.g., 7-AAD (listed below), TO-PRO-3, or DAPI, among many others,
for viability measurements though propidium iodide remains the most
commonly used dye for DNA content analysis.
Texas Red has an excitation maximum in the
yellow-orange range of the color spectrum; consequently, Texas Red
cannot be excited on any of the benchtop analyzers in the core
facility. It can, however, be detected using one of green laser
lines of the MoFlo's argon laser but ideally using the yellow laser
line of the MoFlo's krypton laser. When excited, Texas Red has an
excitation maximum in the orange at 612 nanometers.
7
-Aminoactinomycin D (7-AAD), like propidium
iodide, is a DNA intercalating dye but 7-AAD is specific for C-G
base pairs. It is well suited for viability measurements and also
for apoptosis experiments where it's paired with Annexin V. Unlike
propidium iodide, this dye has minimal emission bleed from the FL3
detector into the FL2 (PE) detector on the FACSCalibur or FACScan.
Whereas PI can be detected in either FL2 or FL3, though it is
typically detected in FL2, of the FACScan or FACSCalibur, 7-AAD is
detected in FL3.
PE:Phycoerythrin 藻红蛋白;488nm,575nm(橙红)
FITC,异硫氰酸荧光素 488nm,525nm(绿
APC:英文全名:allophycocyanin,中文名:别藻青蛋白,最大吸收峰:650nm,最大发射荧光峰:660nm,适用于双激光流式仪,可被600-640nm波长的激光激发。
PerCP英文全名:peridinin chlorophyll
protein,中文名:多甲藻叶绿素蛋白,最大激发波峰490nm,被488nm的氩离子激光激发后,发射光峰值约为677nm,与FITC、PE进行多色荧光染色补偿重叠较少,非特异性结合也较少。虽然较易使用,但量子产量不高,多用于检测表达较高的抗原。